Survey Date | Inspection Type | Track ID | Deficiency | Requirement Description | Correction Date |
11/13/2020 | Fire/Life/Safety | FZBG | K0345 | FIRE ALARM SYSTEM - TESTING AND MAINTENANCE | 01/15/2021 |
11/13/2020 | Fire/Life/Safety | FZBG | K0353 | SPRINKLER SYSTEM - MAINTENANCE AND TESTING | 12/23/2020 |
11/13/2020 | Fire/Life/Safety | FZBG | K1005 | GENERAL EQUIPMENT TESTING & MAINTENANCE | 02/08/2021 |
11/13/2020 | Fire/Life/Safety | FZBG | K1009 | JOINT PENETRATIONS | 02/08/2021 |
11/13/2020 | Fire/Life/Safety | FZBG | K1150 | SECURITY MANAGEMENT | 12/12/2020 |
08/15/2018 | Fire/Life/Safety | 4PQ1 | K0345 | FIRE ALARM SYSTEM - TESTING AND MAINTENANCEFire Alarm System - Testing and Maintenance
A fire alarm system is tested and maintained in accordance with an approved program complying with the requirements of NFPA 70, National Electric Code, and NFPA 72, National Fire Alarm and Signaling Code. Records of system acceptance, maintenance and testing are readily available.
9.6.5, 9.6.7, 9.6.8, and NFPA 70, NFPA 72
| 09/25/2018 |
08/15/2018 | Standard | 4PQ1 | Q0242 | INFECTION CONTROL PROGRAMThe ASC must maintain an ongoing program designed to prevent, control, and investigate infections and communicable diseases. In addition, the infection control and prevent program must include documentation that the ASC has considered, selected, and implemented nationally recognized infection control guidelines.
| 09/25/2018 |
12/11/2017 | Fire/Life/Safety | 5II8 | K0291 | EMERGENCY LIGHTINGEmergency Lighting
Emergency lighting of at least 1-1/2-hour duration is provided automatically in accordance with 7.9.
18.2.9.1, 19.2.9.1
| 02/07/2018 |
12/11/2017 | Fire/Life/Safety | 5II8 | K0345 | FIRE ALARM SYSTEM - TESTING AND MAINTENANCEFire Alarm System - Testing and Maintenance
A fire alarm system is tested and maintained in accordance with an approved program complying with the requirements of NFPA 70, National Electric Code, and NFPA 72, National Fire Alarm and Signaling Code. Records of system acceptance, maintenance and testing are readily available.
9.6.5, 9.6.7, 9.6.8, and NFPA 70, NFPA 72
| 02/07/2018 |
12/11/2017 | Fire/Life/Safety | 5II8 | K0353 | SPRINKLER SYSTEM - MAINTENANCE AND TESTINGSprinkler System - Maintenance and Testing
Automatic sprinkler and standpipe systems are inspected, tested, and maintained in accordance with NFPA 25, Standard for the Inspection, Testing, and Maintaining of Water-based Fire Protection Systems. Records of system design, maintenance, inspection and testing are maintained in a secure location and readily available.
a) Date sprinkler system last checked _____________________
b) Who provided system test ____________________________
c) Water system supply source __________________________
Provide in REMARKS information on coverage for any non-required or partial automatic sprinkler system.
9.7.5, 9.7.7, 9.7.8, and NFPA 25
| 02/07/2018 |
|
|